The LASIK industry & the FDA have conspired since LASIK's inception to purposely withhold information vital to the public in making a truly informed LASIK decision. With Lasikdecision.com, The hope is to show you what the industry and FDA would not and did not even think of doing until LASIK casualties started speaking out, and yet, they still did NOTHING.
Precision of Flap Measurements For LASIK In 4428 Eyes PDF Print E-mail
Monday, 12 December 2005 19:00

1: J Refract Surg. 2003 Mar-Apr;19(2):113-23.

Flanagan GW, Binder PS.

Gordon Binder Vision Institute, San Diego, CA, USA.

PURPOSE: To determine the factor(s) that influence the dimensions and predictability of the LASIK corneal flap with the Automated Corneal Shaper (ACS) or the Summit Krumeich Barraquer microkeratome (SKBM).

METHODS: We performed a retrospective, comparative interventional case study of 4,428 eyes. Flap dimensions were measured using subtraction ultrasonic pachymetry during LASIK with one of two microkeratomes.

RESULTS: Mean preoperative corneal thickness for all eyes was 555 +/- 35 microm. Corneal curvature and refractive astigmatism were inversely related to preoperative corneal thickness (P<.001). With an attempted flap thickness of 160 microm, the ACS flap thickness averaged 119.8 +/- 22.9 microm; SKBM flaps averaged 160.9 +/- 24.1 microm (P<.001). The coefficient of variation for central pachymetry compared to flap thickness was 6.4% vs. 22.1%. Flap thickness at enhancement was 10 to 17 microm thicker than at primary surgery. An increase in flap thickness was associated with thicker preoperative pachymetry (P<.001) and younger age for both instruments (P<.001) whereas increasing flap thickness was related to flatter preoperative mean keratometry for the ACS (P<.001) and steeper mean keratometry for the SKBM (P=.005). Less preoperative hyperopia or more myopia was related to an increase in flap thickness only for the SKBM (P<.001).

CONCLUSIONS: Flap thickness varies significantly depending on the microkeratome used. Factors that influence flap thickness are primarily corneal thickness, patient age, preoperative keratometry, preoperative refraction including astigmatism, and corneal diameter. By understanding the factors that affect flap thickness, one can select a microkeratome system to allow maximum refractive correction while minimizing the risk of ectasia.