The LASIK industry & the FDA have conspired since LASIK's inception to purposely withhold information vital to the public in making a truly informed LASIK decision. With Lasikdecision.com, The hope is to show you what the industry and FDA would not and did not even think of doing until LASIK casualties started speaking out, and yet, they still did NOTHING.
LASIK complication: loss of electricity to the microkeratome PDF Print E-mail
Saturday, 10 December 2005 19:00

Acta Ophthalmol Scand. 2003 Oct;81(5):530-2.

LASIK complication: loss of electricity to the microkeratome during the forward pass.

Tuominen IS, Tervo TM. Department of Ophthalmology, Helsinki University Hospital, PO Box 220, HUS, SF-00029 Helsinki, Finland. ilpo.tuominen@hus.fi

CONTEXT: A 32-year-old woman was scheduled for myopic laser in situ keratomileusis (LASIK) because of myopia and anisometropia caused by retinal detachment surgery.

CASE REPORT: During surgery, a sudden malfunction of the microkeratome during the forward pass was experienced. It was not possible to reverse the blade manually along the suction ring. Moreover, disconnecting the suction from the control unit did not help at first, because the suction ring was firmly attached to the ocular surface. However, detaching the suction line from the control unit aborted the vacuum and allowed the surgeon to turn the whole microkeratome backwards, mimicking the normal blade movement. Finally, an almost normal flap was observed, and the operation was successfully completed. Afterwards, the wire to the electromotor of the microkeratome was found to be broken and subsequently replaced.

CONCLUSION: This type of unforeseen microkeratome malfunction may result in serious flap or other complications.